Orthognathic surgery requires stable osteofixation for effective bone healing and remodeling. Traditionally, titanium plates and screws have been considered the “Gold Standard” for rigid fixation. However, problems associated with titanium plating systems include thermal sensitivity, growth restriction of surrounding bony tissues, interference with radiographic imaging, and screw migration1. Additionally, a report indicated that titanium miniplates are a risk factor for bisphosphonate-related osteonecrosis of the jaw2. As a result, titanium internal fixators often must be removed in a secondary operation.
Biodegradable plates have been introduced as an effective alternative to titanium devices, addressing many of these issues. Since 2000, innovative biodegradable materials have been increasingly utilized in orthognathic surgery3,4 with the advantage of no need for device removal. However, drawbacks of biodegradable plates include weaker mechanical properties compared to titanium alloys and a delayed inflammatory response. Numerous clinical studies have demonstrated comparable stability between biodegradable and titanium osteofixation in orthognathic surgery5,6.
Nevertheless, descriptive information on delayed inflammation following biodegradable osteofixation in orthognathic surgery patients is sparse. This study reviews three patients who underwent biodegradable and osteoconductive osteofixation after orthognathic osteotomy and exhibited delayed inflammatory responses.
A retrospective review of patients with dentofacial deformities operated on by a single surgeon between 2013 and 2020, was conducted. Three patients were identified who underwent bimaxillary orthognathic osteotomy and fixation with unsintered-hydroxyapatite/poly-L lactic acid (u-HA/PLLA) devices, subsequently developing adverse tissue reactions. These lesions were treated with debridement and/or removal of the fixation devices. Histological evaluations were performed using H&E staining, and structural changes of the devices were assessed using scanning electron microscopy (SEM). Written informed consent was obtained from all patients, and this study was approved by the Institutional Review Board (IRB) of Gangneung-Wonju National University Dental Hospital (IRB No. GWNUDH-IRB2024-A007).
A 23-year-old male patient was referred to our department for a surgery-first orthognathic procedure due to bimaxillary protrusion. Maxillomandibular osteotomies were performed for maxillary posterior movement, mandibular setback, and advancement genioplasty. Mini-sized u-HA/PLLA plates were used for maxillary fixation, and u-HA/PLLA mesh (Osteotrans MX; Takiron) was used for mandibular fixation. Twenty months later, a well-defined, painless swelling was noted on both sides of the mandibular vestibules.(Fig. 1. A, 1. D) Debridement and device removal were performed under local anesthesia (Fig. 1. B, 1. E) and prophylactic antibiotics were prescribed. H&E staining demonstrated multi-nucleated giant cells phagocytosing foreign granular materials with inflammatory cell infiltration, indicative of a foreign-body reaction.(Fig. 1. C, 1. F) SEM imaging of the explanted u-HA/PLLA mesh revealed that the device surface was disintegrating and pores had formed.(Fig. 2. A-2. C) At higher magnification, fibroblasts and exosomes were attached to the surface (Fig. 2. D) and fibrin coagulation materials formed bundles, indicating that the organization process was in progress.(Fig. 2. E, 2. F) However, osteoconductive changes were not detected. This patient experienced a second episode of chronic inflammation 34 months after the operation, and a second debridement was performed.(Fig. 3)
A 29-year-old male was referred to our department due to mandibular prognathism. Maxillomandibular osteotomy was performed for maxillary posterior impaction and mandibular setback. Osteofixation was completed using u-HA/PLLA devices according to our protocol6. Twelve months later, a well-defined, painless swelling was observed on the left-side sagittal spliting ramus osteotomy (SSRO) site.(Fig. 4. A) The inflammation was resolved with drainage and anti-inflammatory medication without device removal. H&E staining revealed a chronic inflammatory lesion with plasma cell and small lymphocyte infiltration.(Fig. 4. B)
A 23-year-old male patient visited our department for correction of secondary maxillary hypoplasia due to a unilateral cleft deformity. Cleft orthognathic surgery was performed with maxillary advancement, mandibular setback, and reduction genioplasty. Osteofixation was completed using u-HA/PLLA devices.(Fig. 5. A) Thirty-five months later, the patient complained of diffuse swelling on the left side of the lower chin area.(Fig. 5. B) The lesion was initially treated with repeated drainage and later with removal of the u-HA/PLLA mesh under conscious sedation. H&E staining revealed a pus-discharge lesion with loose stroma and proliferation of polymorphonuclear leukocytes and small capillaries.(Fig. 5. C) The symptoms resolved after removal of the partially resorbed devices on the left side of the chin, with no recurrence over 14 months of follow-up.
This study aimed to evaluate the inflammatory complications of biodegradable osteofixation devices, particularly u-HA/PLLA, in orthognathic surgery. The results indicate that, while u-HA/PLLA fixation devices offer the advantage of avoiding secondary surgery for removal, they are associated with delayed inflammatory reactions, especially in the mandible. This study also explores potential mechanisms and provides clinical recommendations based on the findings.
In 1966, polylactic acid was first proposed for surgical implants in an animal experiment7. Subsequently, three biodegradable materials, polyglycolic acid (PGA), PLLA, and poly-D-lactic acid (PDLA), have been widely used in clinical settings. These materials are metabolized via hydrolysis and are excreted by the lungs as carbon dioxide and water following the citric acid cycle. The rate of biodegradation depends on the crystallinity and hydrophobicity of the hydrolyzed products8. PDLA, an isomeric form of PLLA, shares the same molecular formula but has different spatial orientations of optically active carbon atoms, resulting in different biomechanical and degradation properties. Degradation of PLLA releases crystalline particles that are highly hydrophobic and more resistant to degradation compared to PDLA9. As a result, foreign-body reactions, although not highly irritable to host cells, are frequently reported with the use of PLLA due to its prolonged degradation period10,11.
Over the decades, biodegradable devices have evolved, exhibiting improved degradation rates and decreased adverse tissue reactions by altering the constituent ratios. The polymer ratio scan affects the host’s immunological response to biodegradable implants. Between 2002 and 2012, the author used second-generation bioresorbable poly-L/D-lactide plates consisting of 70% PLLA and 30% PDLA without observing delayed inflammatory responses. This could be attributed to the amorphous structure of the copolymer, which does not release crystalline particles12.
In 1999, hydroxyapatite was physically incorporated into PLLA due to its documented osteoconductive properties. The material used in the study was au-HA/PLLA nanocomposite13. Clinically, the degradation period of u-HA/PLLA is longer than 5.5 years14, potentially leading to foreign-body reactions as a low degradation rate is related with increased metabolite accumulation. At a later stage of u-HA/PLLA biodegradation, the disintegrated PLLA undergoes enzymolysis. In this decomposition stage, the host tissue’s clearance capacity may be overloaded, increasing the risk of adverse tissue reactions15. The remaining crystalline PLLA particles may trigger a foreign-body tissue reaction, and the uneven release of PLLA fragments can provoke a physical inflammatory response.
An ideal biodegradable fixation device should fully resorb once the bone healing process is complete, with resulting metabolites causing no local or systemic reactions. A delayed foreign-body reaction is defined as inflammatory symptoms occurring 6 months after implantation and typically does not involve infection16. In present study, the onset of inflammation occurred 20, 12, and 35 months after the operations, with the spectrum of adverse tissue responses ranging from localized nonspecific inflammation (Case 2) to severe foreign-body reactions necessitating device removal (Cases 1 and 3). The timing of the tissue reaction and the clinical findings of fragmented u-HA/PLLA devices in Cases 1 and 3 indicate a foreign-body reaction during the late degradation process, while the relatively early-stage reaction in Case 2 is notable.
The management of delayed inflammation varied according to the clinical manifestations. Mild-to-moderate inflammation (Case 2) may be treated with drainage and anti-inflammatory drugs. However, severe or recurrent inflammation (Case 1) and/or secondary infections (Case 3) require device removal. In this study, histological findings in Case 1 were consistent with those typically observed in aseptic foreign-body reactions, demonstrating foreign particles surrounded by multinucleated giant cells with inflammatory cell infiltration. The lesion in Case 2 developed 12 months after implantation, likely due to acidic metabolites such as fragmented lactic acid, which can evoke inflammation and recruit inflammatory cells, as seen in the histologic specimen. This phenomenon can lead to the accumulation of sterile fluid, necessitating surgical drainage. The histological findings in Case 3 revealed a suppurative lesion, likely due to secondary inoculation of intraoral pathogens through an intraoral incision. SEM images from Case 1 revealed that the implant was undergoing biodegradation and was well-integrated with the surrounding host tissues.
Factors such as the degradation rate of polymers, implant volume and shape, and recipient environment may affect the occurrence of foreign-body reactions17. A low degradation rate is associated with increased accumulation of metabolites, which can increase the risk of adverse reactions, as described above. If the volume and shape of the devices allow accumulation of sterile fluid from an immunological reaction, the clearance capacity of the host tissue may decrease, leading to clinical symptoms. Considering the shape of the device, the mesh plate used in this study may be more advantageous than other forms of u-HA/PLLA devices due to its even release of metabolites. Most studies reported a less than 5% occurrence rate of delayed inflammation in patients with biodegradable plates. However, no adverse tissue reactions occurred in 44 pediatric patients with facial fractures18, reflecting more favorable recipient environments.
In this study, delayed inflammation occurred only in the mandible and did not occur in the maxillary fixation site, which was fixed with four u-HA/PLLA plates in the standard position. The author suggests that the coarse structure and high vascularity of the maxillary bone may prevent the accumulation of exudate from a foreign-body reaction. However, the rate of delayed foreign-body reactions caused by biodegradable PLLA/PGA or PLDLA/trimethylene carbonate plates is 3.4% among 234 patients with zygomaticomaxillary fractures19. We also suggest that there are differences in the characteristics of patient groups between those undergoing orthognathic surgery and trauma patients. In general, the orthognathic surgery patient group is within an age range that has a stronger immune system compared to the trauma patient group. In this study, the incidence of delayed inflammatory reactions at the SSRO site is less than 1%.
Additionally, the risk of foreign-body reactions appears to be independent of gender20 even though all three patients in this study were male. However, it has been reported that smokers have a higher rate of delayed inflammatory reactions compared to nonsmokers21. The severity of inflammatory responses varies among individuals, likely due to the differences in immune status, including hyper-reaction and sensitivity to degradation products. The patient in Case 1 exhibited bilateral and repeated occurrences of delayed inflammation, and the source of the second inflammation was possibly a residual u-HA/PLLA screw, as seen in Fig. 3. B. It is possible that this patient had a genetic predisposition to sensitivity to PLLA metabolites, which warrants further study.
The author encountered three cases of delayed inflammation following orthognathic osteotomy and u-HA/PLLA osteofixation. The lesions occurred only in the mandible and were treated uneventfully with routine drainage and/or device removal without further complications. It is important to inform orthognathic patients undergoing u-HA/PLLA osteofixation about the possibility of delayed inflammation, and long-term follow-up is recommended.